Displaying report 1-1 of 1.
Reports until 16:30, Sunday 25 September 2016
H1 AOS
sheila.dwyer@LIGO.ORG - posted 16:30, Sunday 25 September 2016 - last comment - 10:25, Wednesday 28 September 2016(29964)
PSL tripped

The PSL tripped around 21:37 UTC, it took us about an hour to recover from this . We called Jason and Peter, Jason called back and walked me through restarting it remotely.  I added about 200mL of water to the chiller, there was water on the floor in front of the chiller.  Jason confirmed that this is normal, and that the turbulence and bubbles in the chiller tube are from a known slow leak. 

Comments related to this report
jason.oberling@LIGO.ORG - 11:10, Monday 26 September 2016 (29974)

PSL tripped due to the power meter flow.  The 1st attachment shows the power meter flow dropping just before the crystal chiller interlock trips.  The 2nd attachment shows the power meter flow dropping ~1 second before the FE flow drops.  Interestingly, it takes ~6 seconds after the power meter flow drops and the crystal chiller interlock trips for the crystal chiller flow to drop (3rd attachment).  Not sure if this is due to the rate that the Beckhoff PC polls the crystal chiller, or if it really takes 6 seconds from interlock trip to chiller shutdown (I suspect the former).  The full timeline of the trip, assuming t0 = 21:37:51 UTC (time of trip):

  • t0:  Power meter flow goes to 0 lpm
  • t0 + <1 sec:  Crystal chiller interlock trip, FE and HPO lose power (due to IL trip)
  • t0 +1 sec: FE and Laser Head flows go to 0 lpm
  • t0 + 6 sec:  Crystal chiller flow goes to 0 lpm
Images attached to this comment
jason.oberling@LIGO.ORG - 10:25, Wednesday 28 September 2016 (30043)

Filed FRS #6319.

Displaying report 1-1 of 1.