Displaying report 1-1 of 1.
Reports until 14:34, Friday 07 October 2016
H1 ISC (ISC)
gabriele.vajente@LIGO.ORG - posted 14:34, Friday 07 October 2016 (30311)
BruCo scans for PRCL, SRCL and MICH error signals

In relation to Sheila's investigation on the increased noise in the aux length loops, I ran BruCO on the loop error signals LSC-PRCL_IN1, LSC-SRCL_IN1, LSC-MICH_IN1.

PRCL

The BruCo report can be found here:

https://ldas-jobs.ligo.caltech.edu/~gabriele.vajente/bruco_prcl_1159781057/

Beside some expected coherences, there are quite some ASC signals that are coherent in the 10 to 100 Hz region. Of course, it might as well be that the angular signals are polluted by the LSC control.

Images attached to this report
Displaying report 1-1 of 1.