H1 PSL
patrick.thomas@LIGO.ORG - posted 21:15, Saturday 08 October 2016 - last comment - 23:18, Saturday 08 October 2016(30339)
PSL tripped?
It appears that the PSL may have tripped after the last lock loss. The NPRO output power is 0.
Comments related to this report
patrick.thomas@LIGO.ORG - 21:24, Saturday 08 October 2016 (30340)
There is a flow sensor error on the top chiller and water on the front of it. The fill cap is still in place.
patrick.thomas@LIGO.ORG - 21:35, Saturday 08 October 2016 (30341)
Talked with Peter. He talked me through turning off the auto locking and laser. Since no one else is here I am going to head out and he will come in tomorrow morning.
kiwamu.izumi@LIGO.ORG - 23:18, Saturday 08 October 2016 (30344)

PSL restarted without no issue. The errors were xtal chiller and heads 1-4 flow error. We added 200 ml water.