It appears that the PSL may have tripped after the last lock loss. The NPRO output power is 0.
There is a flow sensor error on the top chiller and water on the front of it. The fill cap is still in place.
Talked with Peter. He talked me through turning off the auto locking and laser. Since no one else is here I am going to head out and he will come in tomorrow morning.
PSL restarted without no issue. The errors were xtal chiller and heads 1-4 flow error. We added 200 ml water.
PSL restarted without no issue. The errors were xtal chiller and heads 1-4 flow error. We added 200 ml water.