state of H1: IMC was lockednow unlocked
Help: Jason, Peter
Ditails:
-
laser tripped last night
-
I restarted with Jason's help
-
I filled both chillers with about 100ml
-
Peter was on site when I returned to the CR
-
the laser tripped out again
-
Peter restarted and went into the PSL to realign
-
IMC locked about 30 minutes ago and has stayed locked
-
NPRO_NoiseEater is complaining that it's out of range
-
diffracted power is coming up on DIAG_MAIN as HIGH at 6%
-
Peter is going out to fix
-
Robert tripped MC3 with shaking, IMC is out of lock