J. Kissel for S. Karki I've changed the > 1 kHz PCALX calibration line frequency, a.k.a. the "long duration sweep" line, to 2501.3 Hz. Recall this started moving again on Oct 6th (see LHO aLOG 30269) I report the progress towards completing the sweep below. Frequency Planned Amplitude Planned Duration Actual Amplitude Start Time Stop Time Achieved Duration (Hz) (ct) (hh:mm) (ct) (UTC) (UTC) (hh:mm) --------------------------------------------------------------------------------------------------------------------------------------------------------- 1001.3 35k 02:00 1501.3 35k 02:00 2001.3 35k 02:00 2501.3 35k 05:00 39322.0 Oct 12 2016 03:20:41 UTC 3001.3 35k 05:00 39322.0 Oct 06 2016 18:39:26 UTC Oct 12 2016 03:20:41 UTC days 3501.3 35k 05:00 39322.0 Jul 06 2016 18:56:13 UTC Oct 06 2016 18:39:26 UTC months 4001.3 40k 10:00 4301.3 40k 10:00 4501.3 40k 10:00 4801.3 40k 10:00 5001.3 40k 10:00