peter.king@LIGO.ORG - posted 12:47, Thursday 13 October 2016 - last comment - 18:54, Thursday 13 October 2016(30497)
crystal chiller flow rate
Using the crystal chiller that is currently sitting in the mechanical room, the on/off flow rate
was recorded as a function of time.  The data is that reported via the RS-232 interface.  When
switched off, the flow rate drops to zero in 3.245 seconds.  This seems a lot slower than the
drop reported by EPICS, which seems to suggest the flow rate drops to zero in less than a second.
Images attached to this report
Comments related to this report
john.worden@LIGO.ORG - 18:54, Thursday 13 October 2016 (30515)

The difference in times might be a result of different flow impedance - ie -different plumbing and thermal loads connected to the two chillers.