Reports until 13:01, Thursday 13 October 2016
H1 General (GRD, ISC, OpsInfo)
cheryl.vorvick@LIGO.ORG - posted 13:01, Thursday 13 October 2016 - last comment - 14:02, Thursday 13 October 2016(30498)
made it to DC readout - now in NLN

issues getting here:

Comments related to this report
cheryl.vorvick@LIGO.ORG - 13:14, Thursday 13 October 2016 (30500)

 REFL WFS centering not coming on in Inital alignment for PRC and SRC, so in both cases those didn't complete until engaged by hand (guardian?)

Images attached to this comment
cheryl.vorvick@LIGO.ORG - 13:12, Thursday 13 October 2016 (30501)

POPA OFFSETS - Evan looks at these - not sure they're OK but in the end they seem to work

Images attached to this comment
cheryl.vorvick@LIGO.ORG - 14:02, Thursday 13 October 2016 (30503)

    ASC - Evan toggled PRC1_P on and off (3 seconds each) to save the lock and also allow PRC1 P to converge

Images attached to this comment
cheryl.vorvick@LIGO.ORG - 13:17, Thursday 13 October 2016 (30504)

    POPA sum got noisy in ENGAGE SOFT LOOPS until I engaged the CHARD offset of 0.1 by hand in Engage Soft Loops, then the signal improved

Images attached to this comment
cheryl.vorvick@LIGO.ORG - 13:22, Thursday 13 October 2016 (30505)

    I set the ramp on CSOFT to 120 seconds, and at 120 seconds there's a kick that can be seen on CSOFT_P

  • in the picture the offset ramps starting at the skip in data between -3 and -2
  • between -1 and 0, 120 seconds later the ramp ends and there's a glitch
  • isn't this what's not supposed to happen, i.e. a ramp shouldn have a sharp ending?
Images attached to this comment
cheryl.vorvick@LIGO.ORG - 13:24, Thursday 13 October 2016 (30506)

    PI mode 26 rang up and did a bouncing thing until I lowered the gain by half from -5000 to -2500, and then tuned the phase