Displaying report 1-1 of 1.
Reports until 16:07, Saturday 22 October 2016
H1 General
edmond.merilh@LIGO.ORG - posted 16:07, Saturday 22 October 2016 - last comment - 19:02, Saturday 22 October 2016(30751)
Shift Summary -Day
Comments related to this report
terra.hardwick@LIGO.ORG - 16:31, Saturday 22 October 2016 (30753)

My earlier comment assumption was correct: during the settling of the recent ring heater changes, Mode26 had shifted outside of the usual guardian-controlled bandpasses; an already existing filter needed to be turned on. After this, we powered up and damped with no problems. Locked for 2.5 hours at this point needing only normal phase changes. I walked Ed through how to check on this and change the filter, but this large of change only occurs after a ring heater change so will not be a normal issue operators need to worry about. 

Reminder that I've requested to be called anytime there is a persistant PI problem.

terra.hardwick@LIGO.ORG - 19:02, Saturday 22 October 2016 (30755)

I've added step-by-step instructions for how to attempt to handle this scenerio in the Operator PI Wiki (under 'If a PI seems unresponsive') if for some reason I can't be reached. PI Wiki can also be opened from PI medm screen. Working on automating this.

Displaying report 1-1 of 1.