TITLE: 10/22 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Commissioning
INCOMING OPERATOR: Patrick
SHIFT SUMMARY:

Comments related to this report
sheila.dwyer@LIGO.ORG - 14:21, Sunday 23 October 2016 (30771)

I would recomend that instead of putting the extra factor of 2 gain in PRCL2, people double the gain in PRCL 1.  The guardian doesn't touch the gain in PRCL2, but it will later in the locking sequence adjsut PRCL1 to be the nominal value.  If people adjust the gain in PRCL2 and forget to rest it to 1, this can cause problems. 

If we consistently are needing higher gain, there is a parameter in LSC params that adjusts the gains for lock acquisition. 