~0930 hrs. local -> Started heat reduction of X-end RGA
Pumps will continue to run at X-end RGA for a few more days. 

Closed cal-gas isolation valves.  Manually reducing variac's by ~10% every 2 hrs. today.

Bake Summary: 
Approximately 90% of surface area soaked between 180C < temperature < 210C for 96 hrs..  Additional 72 hrs. manual incremental ramp up.  ~10% surface area tapered down to ~90C at turbo inlet line-of-site during soak period.  

. 