Kyle, Chandra We ran all QDP80 roughing pumps on site today for maintenance and for refresher (WP 6269); checked for gross leaks in foreline/turbo volumes. Found one at MX. EY as found pressure foreline = 14 Torr turbo = 0.58 Torr after rough pump down and following valve isolation foreline = 0.040 Torr turbo = 0.036 Torr MY as found pressure foreline = 2.3 Torr turbo = 0.60 Torr after rough pump down and following valve isolation foreline = 0.013 Torr turbo = 0.022 Torr LVEA - vertex as found pressure foreline = 2.2 Torr turbo = 0.67 Torr after rough pump down and following valve isolation foreline = 0.017 Torr turbo = 0.022 Torr LVEA - Xbeam manifold as found pressure foreline = 12 Torr turbo = 0.63 Torr after rough pump down and following valve isolation foreline = 0.021 Torr turbo = 0.039 Torr LVEA - diagonal as found pressure foreline = 5.4 Torr turbo = 0.54 Torr after rough pump down and following valve isolation foreline = 0.026 Torr turbo = 0.039 Torr LVEA - Ybeam manifold as found pressure foreline = 5.4 Torr turbo = 0.48 Torr after rough pump down and following valve isolation foreline = 0.025 Torr turbo = 0.034 Torr MX as found pressure foreline = 780 Torr turbo = 1.0 Torr after rough pump down and following valve isolation foreline = leaking turbo = 0.084 Torr EX as found pressure foreline = 6.6 Torr turbo = 0.43 Torr after rough pump down and following valve isolation foreline = 0.017 Torr turbo = 0.026 Torr The leak in the X-mid fore line is between the Turbo Header isolation valve and the Turbo fore line Safety Valve. Removed and examined the NW seals but nothing obvious. We verified each QDP80 shuts down with the absence of N2 purge seal gas supply except the Y-end QDP80 which did not shut down. Y-mid pump turned off due to overheating because cooling water was not valved in initially. It leaked a small puddle of coolant and ran OK upon cooling. We pumped down the purge air line in LVEA unintentionally.