Displaying report 1-1 of 1.
Reports until 13:24, Monday 31 October 2016
H1 INJ
evan.goetz@LIGO.ORG - posted 13:24, Monday 31 October 2016 - last comment - 13:26, Monday 31 October 2016(31033)
Injection successful after restarting Guardian node
Evan G., Dave B., Chris B., Thomas S.

After completely restarting the INJ_TRANS Guardian node, this finally fixed the transient injection problem as reported in LHO aLOG 31030. The interferometer was not locked, so no point in trying to recover the signal in a low-latancy analysis. This is the entry in the schedule file:
1161980000 INJECT_CBC_ACTIVE 0 1.0 /ligo/home/evan.goetz/lscrepos/Inspiral/H1/coherentbbh7_1126259455_H1.out

I observed the signal in H1:CAL-PINJX_HARDWARE_OUT, H1:CAL-PCALX_RX_PD_VOLTS_OUT, and H1:CAL-PCALX_OFS_AOM_DRIVE_MON_IN2.
Comments related to this report
evan.goetz@LIGO.ORG - 13:26, Monday 31 October 2016 (31034)
Please note the following warning about "Aborting Logging" in the log file:
2016-10-31T20:12:43.18044 INJ_TRANS executing state: INJECT_CBC_ACTIVE (101)
2016-10-31T20:12:43.18068 INJ_TRANS [INJECT_CBC_ACTIVE.enter]
2016-10-31T20:12:43.18399 INJ_TRANS [INJECT_CBC_ACTIVE.main] USERMSG 0: INJECTION ACTIVE: 1161980000.000000
2016-10-31T20:12:51.05950 Aborting Logging because SIStrLog call failed -1
2016-10-31T20:13:19.57595 INJ_TRANS [INJECT_CBC_ACTIVE.main] GPS start time: 1161980000.000000
2016-10-31T20:13:19.57602 INJ_TRANS [INJECT_CBC_ACTIVE.main] Requested state: INJECT_CBC_ACTIVE
Displaying report 1-1 of 1.