Reports until 05:42, Wednesday 02 November 2016
H1 TCS
nutsinee.kijbunchoo@LIGO.ORG - posted 05:42, Wednesday 02 November 2016 - last comment - 21:38, Monday 07 November 2016(31116)
Glitches in CO2 temperature channels

Started yesterday evening. Doesn't seems to be equally spaced.

Images attached to this report
Comments related to this report
alastair.heptonstall@LIGO.ORG - 08:57, Wednesday 02 November 2016 (31124)

It could be pickup through the unshielded part of the RTD cable (where it comes out of the laser), but it's strange that it started in X and Y at the same time.  Can you check the FLIR cameras are turned off on the MEDM screen.

betsy.weaver@LIGO.ORG - 09:07, Wednesday 02 November 2016 (31125)

Apparently there is a "glitch" in the laser power of both X and Y TCS systems every time the Laser power meter sees a request for a power change.  (see 1 day trend with additional channels to correlate with).  However, this did not just start yesterday, it has been going for at least a month (see second plot of the last 3 days).  I spot checked back a month and the behavior is the same.  Note, there are a few extra spikes on these channels but this doesn't seem to be new behavior.

Images attached to this comment
nutsinee.kijbunchoo@LIGO.ORG - 21:38, Monday 07 November 2016 (31299)

Update: I looked closer at these glitches.

 

Characteristic:

A single glitch spans across multiple sample points, lasted a little less than half a second. Almost everyone of these glitches lead by one smaller glitch that's roughly 1.5 second apart, followed by an even smaller glitch and immediately followed by a larger glitch.

 

Timeline:

I looked back several months and it seems like these glitches started to show up much more frequently on May 13th this year (started at 16:28 UTC to be precise). I looked back on the alog and didn't find any invasive work with the TCS on May12th or 13th.

Images attached to this comment