Displaying report 1-1 of 1.
Reports until 00:02, Wednesday 02 November 2016
H1 ISC
stefan.ballmer@LIGO.ORG - posted 00:02, Wednesday 02 November 2016 - last comment - 13:12, Wednesday 02 November 2016(31103)
A2L DRIVEALIGN gains over the last 14 days

Notes:

0) A2L gains of order two are getting suspicious, because it essentially means some coils are shut off.

1) ITMY_Y2L was always large (~-2) , but since 10/29 7UTC it is even bigger.

2) ITMX_Y2L also got big (~2) on 10/29 7UTC

Not sure what to make of this...

Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 13:12, Wednesday 02 November 2016 (31136)DetChar, SUS
Tagging SUS. 

We should check functionality of coils.

EDIT: Checked the functionality of the coils by taking a quick transfer function between COILOUT_EXC channels (OSEM basis excitation channels closest to the DAC output), and the various coil driver monitor channels. Though I've not taken the time to calibrate or understand the signals, I see no differences between the quadrants, and plenty of coherence between all drives and responses.

I don't think there's anything wrong with the ITM L2 actuators.

Note to self -- next time include OSEM sensors and optical levers in your template for those who don't trust the functionality of the coil driver monitor channels.

The templates (attached) live here:
/ligo/home/jeffrey.kissel/2016-11-02/
(apologies for my laziness -- I figured it would be faster to just create new templates than to find the one's Ed used to characterize the monitor signals that are likely committed to the SusSVN. #slapsownwrist)

Also note that 
- The IFO was DOWN during this measurement.
- I turned off the PIT optical lever damping during the measurement so as to not confuse anything.
Images attached to this comment
Non-image files attached to this comment
Displaying report 1-1 of 1.