Reports until 13:11, Wednesday 02 November 2016
H1 ISC
daniel.sigg@LIGO.ORG - posted 13:11, Wednesday 02 November 2016 - last comment - 12:55, Thursday 03 November 2016(31138)
IMC readbacks

Matt Daniel Fil (WP 6294)

The MCL and MCF readbacks have a 10/100 Hz Sallen-Key whitening stage which amplifies the high frequency spectrum to get above ADC noise. Since a while we have observed a 20-50 mHz/√Hz flat noise level in the these spectra when we are locked with the IMC only. Looking with the oscilloscope we estimated about 10m V signal between 100 kHz and 1 MHz, before the whitening. This seems too much for the AA board, so we included additional low pass filters in the readbacks with cut-offs around 15 kHz. A 15/150 kHz pole-zero was added to the Sallen-Key, and another 15 kHz pole was added to the output stage.

In detail (common mode board, IMC, s/n 1102626):

The attached spectra now show a frequency noise level which is compatible with the one observed in full lock. The coherence is also improved. The ADC noise is not too far away in regions with reduced coherence.

Non-image files attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 16:12, Wednesday 02 November 2016 (31142)

Here is a comparison between MCF fully locked and 2W IMC only (REF traces). The changes are much smaller now, indicating that MCF sees frequency noise from the laser.

evan.hall@LIGO.ORG - 12:55, Thursday 03 November 2016 (31166)

The IMC shot noise limit here should be about 1 mHz/Hz1/2, assuming 0.3 mW of light (mostly carrier) on the PD with the IMC locked, 5 mW of light on the PD with the IMC unlocked, and a modulation depth of 0.01 rad.

Non-image files attached to this comment