Reports until 13:58, Thursday 03 November 2016
LHO VE
kyle.ryan@LIGO.ORG - posted 13:58, Thursday 03 November 2016 - last comment - 16:26, Friday 04 November 2016(31170)
60 day trend of PT140
Attached is a 60 day trend of PT140 which is a one of the new Inficon BPG402s?  IP7 and IP8 have been a steady 5000 volts for this time period.  Is this a gauge thing?  I haven't been intimate with what Gerardo, John and Chandra have learned regarding the behavior of these new wide-range Bayard-Alpert/Pirani hybrids but this slope looks "not insignificant"
Non-image files attached to this report
Comments related to this report
chandra.romel@LIGO.ORG - 18:58, Thursday 03 November 2016 (31184)
That slope looks really fishy. Are both IPs fully pumping? What does HAM6 pressure look like (also hot cathode ion gauge)? Did PT 170 and 180 flatten out after degassing?
gerardo.moreno@LIGO.ORG - 13:50, Friday 04 November 2016 (31208)

We think that the pressure increase is due to temperature, see attached.  aLOG noting temperature change.

Images attached to this comment
gerardo.moreno@LIGO.ORG - 16:26, Friday 04 November 2016 (31215)SUS

Since we are talking temperature change in the LVEA, note the vertical change on some of the optics (BS and ITMs), other are affected as well.

Images attached to this comment