corey.gray@LIGO.ORG - posted 09:43, Sunday 06 November 2016 - last comment - 10:31, Sunday 06 November 2016(31256)
3rd Lock Attempt So Far....Knock On Wood

For this current lock (15min so far):

Comments related to this report
corey.gray@LIGO.ORG - 10:31, Sunday 06 November 2016 (31257)

After 20 minutes, we dropped out of lock.  (ITMy did not oscillate after.)

Seconds prior to the lockloss:

  • VERBAL noted that PI Mode 27 was ringing up---& it was ringing up super-fast.  
  • Another possibly related effect was the the SRC 1 & 2 Pitch CONTROL signals were both trending up a few minutes before the PI Mode 27 started its ring up.
  • See attached StripTool!

Will attempt another lock with the ITM Ring Heaters OFF from the get-go.

Images attached to this comment
Displaying report 1-1 of 1.