Displaying report 1-1 of 1.
Reports until 09:09, Monday 07 November 2016
H1 PSL (PSL)
patrick.thomas@LIGO.ORG - posted 09:09, Monday 07 November 2016 (31276)
PSL Weekly Report
Laser Status:
SysStat is good
Front End Power is 34.64W (should be around 30 W)
Front End Watch is GREEN
HPO Watch is GREEN
PMC:
It has been locked 1.0 days, 11.0 hr 50.0 minutes (should be days/weeks)
Reflected power is 26.42Watts and PowerSum = 125.8Watts.

FSS:
It has been locked for 0.0 days 10.0 hr and 37.0 min (should be days/weeks)
TPD[V] = 3.567V (min 0.9V)

ISS:
The diffracted power is around 3.384% (should be 5-9%)
Last saturation event was 0.0 days 10.0 hours and 37.0 minutes ago (should be days/weeks)

Possible Issues:
PMC reflected power is high