H1 CAL (CAL)
travis.sadecki@LIGO.ORG - posted 16:37, Tuesday 08 November 2016 (31338)
New End Station PCal Calibrations

Today we took calibration measurements of both end station PCals.  See full reports attached if interested.  In summary, the newest calibration for EX looks good with the new RX/WS and TX/WS ratios being less than a tenth of a percent from the old values.  For EY, the ratios are ~0.5% from the old values.  Looking at the raw plots, there is some biased noise showing up from the WS and a very suspect shelf that comes and goes during separate measurements.  We would like to retake the EY measurements at the next availability since Darkhan and I believe that it is possibly due to a funky power source we have been using with the WS current amplifier (a 3-way extension cord with a power strip plugged in). 