Displaying report 1-1 of 1.
Reports until 23:36, Monday 14 November 2016
H1 ISC
terra.hardwick@LIGO.ORG - posted 23:36, Monday 14 November 2016 (31483)
30 W PI modes

At 30 (or 32) W we're actively damping PI modes 3, 26, 27, 28. I spent some time the last two days with damping turned off one at a time to get the natural ring ups of each. Rough values below (got to see one ringup each for modes 3, 27, 28, two for 26).

Mode # Freq Optic tau
3 15606 Hz ITMX 0.068 s
26 15009 Hz ETMY 0.092 s
27 47495 Hz ETMY 0.024 s
28 47477 Hz ETMY 0.022 s

I also ran Dr. Evan's circulating-power-from-HEPI script on a recent power up and am getting an estimated circulating arm power of 133.67 kW for 30 W PSL power, though the HEPI displacement should be recalibrated. The usual simple calculation of 30 W (input power) * 0.88 (IMC and faraday) *  0.5 (50/50 BS) * 40 W/W (PRG) * 283 W/W (arm build up) = 149.4 kW. 

Displaying report 1-1 of 1.