Reports until 14:46, Wednesday 16 November 2016
H1 CAL
jeffrey.kissel@LIGO.ORG - posted 14:46, Wednesday 16 November 2016 (31546)
PCALX Roaming Calibration Line Frequency Changed to 4301.3
After a week's worth of struggle with the h1oaf computer, and alignment loop instabilities, we've been blessed with another 16 hour lock stretch. As such, we've got sufficient amount of data for the roaming calibration line to move it from 4001.3 Hz to 4301.3 Hz. 

Though the line has been at 4001.3 Hz for 5 days, I recommend only using data from the two long lock stretches in that time from Nov 12 2016 05:00:00 UTC to Nov 12 2016 21:00:00 UTC and Nov 16 2016 06:00:00 UTC to Nov 16 2016 22:00:00 UTC. However, note that the mean laser power is different from the one lock stretch to another -- 29.5 [W] for the Nov 12 stretch, and 31.1 [W] for the Nov 16th stretch.

Current Schedule Status:
Frequency    Planned Amplitude        Planned Duration      Actual Amplitude    Start Time                 Stop Time                    Achieved Duration
(Hz)         (ct)                     (hh:mm)                   (ct)               (UTC)                    (UTC)                         (hh:mm)
---------------------------------------------------------------------------------------------------------------------------------------------------------
1001.3       35k                      02:00                   39322.0           Nov 11 2016 21:37:50 UTC    Nov 12 2016 03:28:21 UTC      ~several hours @ 25 W
1501.3       35k                      02:00                   39322.0           Oct 24 2016 15:26:57 UTC    Oct 31 2016 15:44:29 UTC      ~week @ 25 W
2001.3       35k                      02:00                   39322.0           Oct 17 2016 21:22:03 UTC    Oct 24 2016 15:26:57 UTC      several days (at both 50W and 25 W)
2501.3       35k                      05:00                   39322.0           Oct 12 2016 03:20:41 UTC    Oct 17 2016 21:22:03 UTC      days     @ 50 W
3001.3       35k                      05:00                   39322.0           Oct 06 2016 18:39:26 UTC    Oct 12 2016 03:20:41 UTC      days     @ 50 W
3501.3       35k                      05:00                   39322.0           Jul 06 2016 18:56:13 UTC    Oct 06 2016 18:39:26 UTC      months   @ 50 W
4001.3       40k                      10:00                   39322.0           Nov 12 2016 03:28:21 UTC    Nov 16 2016 22:17:29 UTC      days     @ 30 W
4301.3       40k                      10:00                   39322.0           Nov 16 2016 22:17:29 UTC                                  
4501.3       40k                      10:00
4801.3       40k                      10:00  
5001.3       40k                      10:00

Images attached to this report