State of H1: locked 16+ hours
Commissioners: Sheila
Activities:
-
1:25UTC - H1 out of Observe
-
Sheila took PR3 from POP WFS to REFL WFS, ran a2l, ran other measurements
-
3:01UTC - H1 in Observe
-
4:35UTC - H1 out of Observe
-
I damped ETMY violin modes 1 and 7
-
4:59UTC - H1 in Observe
Note from Sheila:
-
new state PR3_POP_WFS, not in graph, but reachable in Manual, and gives us a way to revert to POP WFS from REFL WFS
Gains:
-
PI mode26 gain = 3000
-
PI mode27 gain = 5000
-
PI mode28 gain = -3000
-
all other PI mode gains at zero
-
ETMY violin mode1 gain = -10
-
ETMY violin mode7 gain = -60
-