H1 CAL (CAL)
aaron.viets@LIGO.ORG - posted 17:36, Monday 28 November 2016 (31941)
Comparison of calibration factors from CALCS, SLM, and GDS
I have plotted about 18 minutes of data starting at GPS time 1164290368 (2016/11/27, 13:59:11 UTC). The attached pdf file has a time series plot of the kappas and the coherences on the first page, and histograms with averages and standard deviations on the second page. The output of CALCS, SLM, and GDS are included. For kappa_tst and kappa_pu, CALCS and GDS are ~1% different, while SLM values are ~4% larger. Agreement is fairly good for kappa_c and f_c, with the exception that the CALCS value of kappa_c is ~1-2% larger.