Reports until 20:00, Friday 02 December 2016
H1 General (SEI)
edmond.merilh@LIGO.ORG - posted 20:00, Friday 02 December 2016 (32130)
Mid-Shift Summary - Eve

 

 

01:06UTC I reverted back to the original sei configuration for BSCs and locking attempts continue.

01:34UTC multiple locklosses at FIND_IR. Reverting back to new configs just to see if I can get past this stage.

01:50UTC New sei configs got me into DRMI land. MICH alignment looks pretty terrible. Trending witness sensors not revealing much. Going to try some alignment with INITIAL_ALIGNMENT guardian. Not going to do the arms if I can help it.

02:00UTC INITIAL_ALIGNMENT/INPUT_ALIGN not working. After trying to adjust the gain higher to get a lock I decided to trend(TimeMachine)IM4. I found it to be about 100urads out in pit. This seems to have corrected the issue. Perhaps this less-than-optimal pointing has been the sore spot in my locking efforts? Also had to put SR2 back to center before aligning SRC.

04:00UTC H1 Locked, Observing Range is ~72Mpc. Lock stretch so far: 01:08 coincidental with L1.

Images attached to this report