Reports until 17:38, Monday 05 December 2016
LHO VE
chandra.romel@LIGO.ORG - posted 17:38, Monday 05 December 2016 - last comment - 05:33, Tuesday 06 December 2016(32214)
corner pressure

PT170 and PT180 were installed in March 2016 when the X/Y beam manifold volumes were vented. PT140 was installed in Sept. 2016. Attached is trend from time of installation. Looks like PT170 & PT180 have been drifting up since after closing and re-opening isolation GVs 5/7 for HAM6 vent. We burped in accumulated gas from gate annuli but I'm not sure what the second hump is from April 9th. No log entry found.

Images attached to this report
Comments related to this report
chandra.romel@LIGO.ORG - 05:33, Tuesday 06 December 2016 (32235)
Could be water accumulation on gauge ion grid is causing drift.