Displaying report 1-1 of 1.
Reports until 13:16, Tuesday 06 December 2016
H1 CAL (DetChar)
jeffrey.kissel@LIGO.ORG - posted 13:16, Tuesday 06 December 2016 (32253)
PCALX Roaming Calibration Line Frequency Changed from 3501.3 to 4001.3
J. Kissel

Continuing the schedule of roaming PCALX line. We'll run this line for one more day, so that we get two points at 4001.3 Hz with a 30 [W] IFO to check consistency between the two repeats of the schedule, and then shut off the line until further notice.


Frequency    Planned Amplitude        Planned Duration      Actual Amplitude    Start Time                 Stop Time                    Achieved Duration
(Hz)         (ct)                     (hh:mm)                   (ct)               (UTC)                    (UTC)                         (hh:mm)
---------------------------------------------------------------------------------------------------------------------------------------------------------
1001.3       35k                      02:00                   39322.0           Nov 28 2016 17:20:44 UTC  Nov 30 2016 17:16:00 UTC         days    @ 30 W  
1501.3       35k                      02:00                   39322.0           Nov 30 2016 17:27:00 UTC  Nov 30 2016 19:36:00 UTC         02:09   @ 30 W         
2001.3       35k                      02:00                   39322.0           Nov 30 2016 19:36:00 UTC  Nov 30 2016 22:07:00 UTC         02:31   @ 30 W
2501.3       35k                      05:00                   39322.0           Nov 30 2016 22:08:00 UTC  Dec 02 2016 20:16:00 UTC         days    @ 30 W
3001.3       35k                      05:00                   39322.0           Dec 02 2016 20:17:00 UTC  Dec 05 2016 16:58:57 UTC         days    @ 30 W
3501.3       35k                      05:00                   39322.0           Dec 05 2016 16:58:57 UTC  Dec 06 2016 21:09:56 UTC         ~15:00  @ 30 W
4001.3       40k                      10:00                   39322.0           Dec 06 2016 21:09:56 UTC
4301.3       40k                      10:00                   39322.0                
4501.3       40k                      10:00                   39322.0           
4801.3       40k                      10:00                   39222.0           
5001.3       40k                      10:00                   39222.0         

Frequency    Planned Amplitude        Planned Duration      Actual Amplitude    Start Time                 Stop Time                    Achieved Duration
(Hz)         (ct)                     (hh:mm)                   (ct)               (UTC)                    (UTC)                         (hh:mm)
---------------------------------------------------------------------------------------------------------------------------------------------------------
1001.3       35k                      02:00                   39322.0           Nov 11 2016 21:37:50 UTC    Nov 12 2016 03:28:21 UTC      ~several hours @ 25 W
1501.3       35k                      02:00                   39322.0           Oct 24 2016 15:26:57 UTC    Oct 31 2016 15:44:29 UTC      ~week @ 25 W
2001.3       35k                      02:00                   39322.0           Oct 17 2016 21:22:03 UTC    Oct 24 2016 15:26:57 UTC      several days (at both 50W and 25 W)
2501.3       35k                      05:00                   39322.0           Oct 12 2016 03:20:41 UTC    Oct 17 2016 21:22:03 UTC      days     @ 50 W
3001.3       35k                      05:00                   39322.0           Oct 06 2016 18:39:26 UTC    Oct 12 2016 03:20:41 UTC      days     @ 50 W
3501.3       35k                      05:00                   39322.0           Jul 06 2016 18:56:13 UTC    Oct 06 2016 18:39:26 UTC      months   @ 50 W
4001.3       40k                      10:00                   39322.0           Nov 12 2016 03:28:21 UTC    Nov 16 2016 22:17:29 UTC      days     @ 30 W (see LHO aLOG 31546 for caveats)
4301.3       40k                      10:00                   39322.0           Nov 16 2016 22:17:29 UTC    Nov 18 2016 17:08:49 UTC      days     @ 30 W          
4501.3       40k                      10:00                   39322.0           Nov 18 2016 17:08:49 UTC    Nov 20 2016 16:54:32 UTC      days     @ 30 W (see LHO aLOG 31610 for caveats)   
4801.3       40k                      10:00                   39222.0           Nov 20 2016 16:54:32 UTC    Nov 22 2016 23:56:06 UTC      days     @ 30 W
5001.3       40k                      10:00                   39222.0           Nov 22 2016 23:56:06 UTC    Nov 28 2016 17:20:44 UTC      days     @ 30 W (line was OFF and ON for Hardware INJ)  
Images attached to this report
Displaying report 1-1 of 1.