H2 Input Mode Cleaner Annulus Ion Pump Installation

(Kyle, Gerardo)

Installed annulus ion pump hardware at the H2 input mode cleaner tube.  The entire system is ready to be pump down and turned on next maintenance day.

Notes about installation: we vented, the already vented annulus space, installed all hardware, pump, controller and managed to pump it down to 4.2x10^-05 torr, then maintenance period was over.

Items to complete next maintenance day: pump the system down, leak check and turn ion pump on.