Displaying report 1-1 of 1.
Reports until 11:02, Wednesday 07 December 2016
H1 CAL (DetChar, OpsInfo)
jeffrey.kissel@LIGO.ORG - posted 11:02, Wednesday 07 December 2016 - last comment - 09:14, Thursday 22 December 2016(32307)
PCALX Roaming Calibration Line Frequency Changed to 4001.3 to 4301.3 Hz
J. Kissel for S. Karki
WP #6368

I've moved the PCALX roaming line, its long duration sweep, from 4001.3 Hz to 4301.3 Hz. See schedule status below.

Since it looks like we'll have the duty cycle to just finish the schedule within a week, I'm just going to continue to re-gathering the highest frequency data. As such, Sudarshan has something consistent and straightforward with which to work instead of data with a bunch details or qualifications. 


Frequency    Planned Amplitude        Planned Duration      Actual Amplitude    Start Time                 Stop Time                    Achieved Duration
(Hz)         (ct)                     (hh:mm)                   (ct)               (UTC)                    (UTC)                         (hh:mm)
---------------------------------------------------------------------------------------------------------------------------------------------------------
1001.3       35k                      02:00                   39322.0           Nov 28 2016 17:20:44 UTC  Nov 30 2016 17:16:00 UTC         days    @ 30 W  
1501.3       35k                      02:00                   39322.0           Nov 30 2016 17:27:00 UTC  Nov 30 2016 19:36:00 UTC         02:09   @ 30 W         
2001.3       35k                      02:00                   39322.0           Nov 30 2016 19:36:00 UTC  Nov 30 2016 22:07:00 UTC         02:31   @ 30 W
2501.3       35k                      05:00                   39322.0           Nov 30 2016 22:08:00 UTC  Dec 02 2016 20:16:00 UTC         days    @ 30 W
3001.3       35k                      05:00                   39322.0           Dec 02 2016 20:17:00 UTC  Dec 05 2016 16:58:57 UTC         days    @ 30 W
3501.3       35k                      05:00                   39322.0           Dec 05 2016 16:58:57 UTC  Dec 06 2016 21:09:56 UTC         ~15:00  @ 30 W
4001.3       40k                      10:00                   39322.0           Dec 06 2016 21:09:56 UTC  Dec 07 2016 18:50:09 UTC         ~20:00  @ 30 W
4301.3       40k                      10:00                   39322.0           Dec 07 2016 18:50:09 UTC
4501.3       40k                      10:00                   39322.0           
4801.3       40k                      10:00                   39222.0           
5001.3       40k                      10:00                   39222.0         

Frequency    Planned Amplitude        Planned Duration      Actual Amplitude    Start Time                 Stop Time                    Achieved Duration
(Hz)         (ct)                     (hh:mm)                   (ct)               (UTC)                    (UTC)                         (hh:mm)
---------------------------------------------------------------------------------------------------------------------------------------------------------
1001.3       35k                      02:00                   39322.0           Nov 11 2016 21:37:50 UTC    Nov 12 2016 03:28:21 UTC      ~several hours @ 25 W
1501.3       35k                      02:00                   39322.0           Oct 24 2016 15:26:57 UTC    Oct 31 2016 15:44:29 UTC      ~week @ 25 W
2001.3       35k                      02:00                   39322.0           Oct 17 2016 21:22:03 UTC    Oct 24 2016 15:26:57 UTC      several days (at both 50W and 25 W)
2501.3       35k                      05:00                   39322.0           Oct 12 2016 03:20:41 UTC    Oct 17 2016 21:22:03 UTC      days     @ 50 W
3001.3       35k                      05:00                   39322.0           Oct 06 2016 18:39:26 UTC    Oct 12 2016 03:20:41 UTC      days     @ 50 W
3501.3       35k                      05:00                   39322.0           Jul 06 2016 18:56:13 UTC    Oct 06 2016 18:39:26 UTC      months   @ 50 W
4001.3       40k                      10:00                   39322.0           Nov 12 2016 03:28:21 UTC    Nov 16 2016 22:17:29 UTC      days     @ 30 W (see LHO aLOG 31546 for caveats)
4301.3       40k                      10:00                   39322.0           Nov 16 2016 22:17:29 UTC    Nov 18 2016 17:08:49 UTC      days     @ 30 W          
4501.3       40k                      10:00                   39322.0           Nov 18 2016 17:08:49 UTC    Nov 20 2016 16:54:32 UTC      days     @ 30 W (see LHO aLOG 31610 for caveats)   
4801.3       40k                      10:00                   39222.0           Nov 20 2016 16:54:32 UTC    Nov 22 2016 23:56:06 UTC      days     @ 30 W
5001.3       40k                      10:00                   39222.0           Nov 22 2016 23:56:06 UTC    Nov 28 2016 17:20:44 UTC      days     @ 30 W (line was OFF and ON for Hardware INJ)  
Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 09:05, Thursday 08 December 2016 (32360)
Note, PCALX was briefly turned OFF during this 4301.3 Hz stretch. See LHO aLOG 32323.
sudarshan.karki@LIGO.ORG - 10:25, Tuesday 13 December 2016 (32512)

Progress on Analysis:

The Pcal to DARM ratio for these individual lines were calculated using SLM Tool and the result is tabulated below. The Cavity pole and the optical gain on the last column are eyeballed measurement from the CalMon Tool . These time dependent corrections havenot been applied to the sensing function analysis yet.

Freq      Start Time               End Time            FFT       Real       Complex       Coh       CC Pole (O. Gain)

---------------------------------------------------------------------------------------------------------------------

1001.3  30-Nov-2016 09:16:44    30-Nov-2016 13:47:04    10       0.022149   -0.358814   0.999589    346.05(0.9863)

---------------------------------------------------------------------------------------------------------------------

1501.3  30-Nov-2016 17:28:11    30-Nov-2016 19:36:18    10      -0.017951   -0.109817   0.999943    346.05(0.9863)

---------------------------------------------------------------------------------------------------------------------

2001.3  30-Nov-2016 19:36:53    30-Nov-2016 21:01:35    10      -0.015679   -0.044511   0.999391    346.05(0.9863)

---------------------------------------------------------------------------------------------------------------------

2501.3  02-Dec-2016 07:52:51    02-Dec-2016 20:16:47    30      -0.011385   -0.020594   0.998321    347.69(0.9703)

---------------------------------------------------------------------------------------------------------------------

3001.3  04-Dec-2016 02:27:58    05-Dec-2016 00:00:17    30      -0.007719   -0.010563   0.999057    347.06(0.9614)

---------------------------------------------------------------------------------------------------------------------

3501.3  06-Dec-2016 02:27:26    06-Dec-2016 06:10:11    30      -0.005708   -0.005431   0.988492    350.70(0.9793)

3501.3  06-Dec-2016 10:36:18    06-Dec-2016 15:03:35    30                                          350.70(0.9793)

---------------------------------------------------------------------------------------------------------------------

4001.3  12-Nov-2016 17:20:18    12-Nov-2016 20:49:17    60      -0.004374   -0.003211   0.996350    350.00(0.9966)

4001.3  12-Nov-2016 09:48:25    12-Nov-2016 16:40:53    60                                          350.00(0.9966)

---------------------------------------------------------------------------------------------------------------------

4301.3  18-Nov-2016 05:20:00    18-Nov-2016 12:44:49    60      -0.003607   -0.002212   0.985922    345.04(0.9748)

4301.3  17-Nov-2016 14:26:40    17-Nov-2016 17:20:13    60                                          347.00(0.9730)

4301.3  17-Nov-2016 08:56:40    17-Nov-2016 11:07:35    60                                          347.00(0.9730)

---------------------------------------------------------------------------------------------------------------------

4501.3  20-Nov-2016 05:00:00    20-Nov-2016 13:24:30    60      -0.003366   -0.002162   0.990728    350.76(0.9985)

4501.3  19-Nov-2016 15:40:00    19-Nov-2016 23:48:16    60                                          350.21(1.0017)

---------------------------------------------------------------------------------------------------------------------

4801.3  22-Nov-2016 07:43:20    22-Nov-2016 16:01:44    60      -0.002883   -0.001369   0.959987    346.89(0.9890)

4801.3  21-Nov-2016 15:20:00    21-Nov-2016 18:22:20    60                                          346.71(0.9979)

4801.3  21-Nov-2016 04:36:40    21-Nov-2016 09:28:48    60                                          346.71(0.9979)

---------------------------------------------------------------------------------------------------------------------

5001.3  25-Nov-2016 07:43:17    25-Nov-2016 14:52:54    60      -0.002634   -0.000904   0.973928    348.94(0.9625)

5001.3  27-Nov-2016 11:44:40    27-Nov-2016 18:04:12    60                                          345.48(0.9724)

---------------------------------------------------------------------------------------------------------------------

 
Data Files:
aligocalibration/trunk/Runs/ER10/H1/Measurements/DARMOLGTFs/HF_2016-11-12_H1_DARM_OLGTF_A_DARMIN2_B_DARMIN1_coh.txt
aligocalibration/trunk/Runs/ER10/H1/Measurements/DARMOLGTFs/HF_2016-11-12_H1_DARM_OLGTF_A_DARMIN2_B_DARMIN1_tf.txt
aligocalibration/trunk/Runs/ER10/H1/Measurements/PCAL/HF_2016-11-12_H1_PCAL2DARMTF_A_PCALYRX_B_DARMIN1_tf.txt
aligocalibration/trunk/Runs/ER10/H1/Measurements/PCAL/HF_2016-11-12_H1_PCAL2DARMTF_A_PCALYRX_B_DARMIN1_coh.txt
 
Parameter Files:
aligocalibration/trunk/Runs/ER10/H1/params/HF_2016-11-12/
 
Script:
aligocalibration/trunk/Runs/ER10/H1/Scripts/SensingFunction/compareSensingHF.m
Non-image files attached to this comment
sudarshan.karki@LIGO.ORG - 09:14, Thursday 22 December 2016 (32833)

More detailed Analysis.

Non-image files attached to this comment