Displaying report 1-1 of 1.
Reports until 06:15, Monday 19 December 2016
H1 General
edmond.merilh@LIGO.ORG - posted 06:15, Monday 19 December 2016 (32728)
PSL Report - Weekly FAMIS #7417
 
Laser Status:
SysStat is good
Front End Power is 34.19W (should be around 30 W)
Front End Watch is GREEN
HPO Watch is GREEN
PMC:
It has been locked 1.0 days, 12.0 hr 51.0 minutes (should be days/weeks)
Reflected power is 14.65Watts and PowerSum = 72.23Watts.
 
FSS:
It has been locked for 0.0 days 9.0 hr and 22.0 min (should be days/weeks)
TPD[V] = 1.314V (min 0.9V)
 
ISS:
The diffracted power is around 4.2% (should be 3-5%)
Last saturation event was 0.0 days 9.0 hours and 21.0 minutes ago (should be days/weeks)
 
Possible Issues:
(Be sure to look into these, as they will not be printed in the report)
PMC reflected power is high
Displaying report 1-1 of 1.