Reports until 06:48, Tuesday 20 December 2016
H1 CAL (CAL)
aaron.viets@LIGO.ORG - posted 06:48, Tuesday 20 December 2016 - last comment - 14:45, Wednesday 04 January 2017(32752)
DCS filters file for offline calibration
Evan G., Aaron V.
I have checked a new filters file into the calibration SVN for offline (C01) calibration. The filters were made using revision #3987 using the script 
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/Scripts/TDfilters/H1_run_td_filters_1165799036.m
The file can be found here:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1165799036.npz

In order to achieve agreement with GDS calibration, it was necessary to add a delay of 2 16kHz clock cycles to the actuation filters. Evan G. is reviewing the DARM model codes to determine where this timing discrepancy originated. 

The attached plots include:
1) h(t) spectrum from CALCS and DCS
2) ASD ratio (DCS / CALCS)
3 - 5) plots of filters and errors
Images attached to this report
Non-image files attached to this report
Comments related to this report
aaron.viets@LIGO.ORG - 14:45, Wednesday 04 January 2017 (32967)
These filters have been removed from the SVN. They were replaced by those discussed in https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=32965
For information on why this was necessary see https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=32907