Reports until 12:20, Tuesday 20 December 2016
H1 TCS
betsy.weaver@LIGO.ORG - posted 12:20, Tuesday 20 December 2016 - last comment - 15:25, Tuesday 03 January 2017(32776)
TCSY In-line Flow Sensor replaced

This morning, Jason, Mark and I swapped the assumed-to-be failing TCSY flow sensor which has been showing epochs of glitching and low readout (while other indicators show normal flow, alogs 32712 and 32230).  The process to do this was such:

 

1) Key laser off at control box in rack, LVEA

2) Turn RF off at mezzanine rack, Mech room

3) Turn chiller off on mezzanine, Mech room

4) Turn power off on back of controller box in rack, LVEA (we also pulled the power cable to the sensor off the front of the controller, but it was probably overkill)

5) Close in-line valves under BSC chamber near yellow sensor to-be-swapped, LVEA

6) Quick-disconnect water tubes at manifold near table, LVEA

7) Pulled yelow top off of yellow sensor housing under BSC at the piping, LVEA

8) Pulled the blue and black wires to the Power recepticles inside the housing (see pic attached).  Pulled full grey cable out of housing.

9) While carefully supporting blue piping*, unscrewed large white nut holding housing/sensor to piping (was tough, in fact so tough that we later removed all of the teflon tape which was unneeded in his join)

10) Pull* straight up on the housing (hard) and it comes out of the piping.

11) Reverse all above steps to insert new housing/sensor, wires and turn everything back on.  Watch for rolled o-rings on the housing and proper alignment of the noth feature when installing the new sensor.  Verify mechanical flow sensors in piping line show ~3-4 G/m readout when flow/chiller is restored to functionality.

12) Setup new flow sensor head with Settings:  Go to the other in-use sensor, pull off the top and scroll through the menu items (red and white buttons on the unit (shown in pic).  Set the new head to these values.

13) Verify the new settings on the head are showing a ~3 G/m readout on the medm screen.  If not, possibly there is setting on the sensor that needs revisited.

14) Monitor TCS to see that laser comes back up and stabilizes.

* Blue piping can crack so be careful to always support it and avoid torque torque

 

Note - with the sensor removed, we could see alot of green merk in the blue piping where the paddle wheel sits.  Still suffering green sludge in this system...

Images attached to this report
Comments related to this report
peter.king@LIGO.ORG - 12:57, Tuesday 20 December 2016 (32777)
A few pictures to add to those already posted.

The O-ring closest to the paddle wheel had a cut to it.  Not near the electronics,
plus there's the other O-ring so it doesn't look like water was getting into where
the electronics is housed.

Some kind of stuff stuck to each blade (paddle?) of the paddle wheel.  Not a good
sign if the cooling water for the laser is meant to be clean.
Images attached to this comment
marc.pirello@LIGO.ORG - 13:20, Tuesday 20 December 2016 (32778)

Settings were as follows:

FLO Unit (Flow Unit) = G/m (default was L/m)

FActor (K-Factor) = 135.00 (default was 20)

AVErage (Average) = 0

SEnSit (Sensitivity) = 0

4 Set (4mA Set Point) = 0 G/m

20 Set (20mA Set Point = 10 G/m (default was 160)

ContrAST (Contrast) = 3

betsy.weaver@LIGO.ORG - 14:05, Tuesday 20 December 2016 (32782)

Here's both TCS system laser power and flow for the past day.  The drop out in the ITMY data is our few hour sensor replacement work.  So far no glitching or low droops.  Although, there weren't any for the last 24 hours on the old sensor either.

Images attached to this comment
jason.oberling@LIGO.ORG - 15:17, Tuesday 03 January 2017 (32931)

Attached is a 14 day duration minute trend of the TCSy chiller flow rate and CO2 laser power since our swap of tthe TCSy flow sensor.  There have been 7 glitches below 2 GPM, with 3 of those glitches being below 1 GPM; all 7 glitches occured in the last week.  Unless the spare flow sensor is also faulty (not beyond belief, but still a hard one to swallow) the root cause of our TCSy flow glitches lies elsewhere.

Images attached to this comment
alastair.heptonstall@LIGO.ORG - 15:25, Tuesday 03 January 2017 (32932)

It might be a good idea to try swapping the laser controller chassis next.  The electronics path for this flow meter is very simple - just the controller and then into the EtherCAT chassis where it's read by an ADC.