Starting CP3 fill. LLCV enabled. LLCV set to manual control. LLCV set to 50% open. Fill completed in 1476 seconds. TC B did not register fill. LLCV set back to 19.0% open. Starting CP4 fill. LLCV enabled. LLCV set to manual control. LLCV set to 70% open. Fill completed in 36 seconds. TC B did not register fill. LLCV set back to 34.0% open.
our reporting script now shows both cp3 and cp4 fills. We will change the format of the text slightly to show the start/stop times because the reporting is delayed beyond the time-out period of one hour.
For CP3:
Increased the LLCV to 20 from 19, due to amount of time it took to overfill CP3 (1476 seconds).