Displaying report 1-1 of 1.
Reports until 12:48, Wednesday 21 December 2016
LHO VE
logbook/robot/script0.cds.ligo-wa.caltech.edu@LIGO.ORG - posted 12:48, Wednesday 21 December 2016 - last comment - 16:30, Wednesday 21 December 2016(32814)
CP3, CP4 Autofill 2016_12_21
Starting CP3 fill. LLCV enabled. LLCV set to manual control. LLCV set to 50% open. Fill completed in 1476 seconds. TC B did not register fill. LLCV set back to 19.0% open.
Starting CP4 fill. LLCV enabled. LLCV set to manual control. LLCV set to 70% open. Fill completed in 36 seconds. TC B did not register fill. LLCV set back to 34.0% open.
Comments related to this report
david.barker@LIGO.ORG - 12:50, Wednesday 21 December 2016 (32815)

our reporting script now shows both cp3 and cp4 fills. We will change the format of the text slightly to show the start/stop times because the reporting is delayed beyond the time-out period of one hour.

gerardo.moreno@LIGO.ORG - 14:59, Wednesday 21 December 2016 (32818)VE

For CP3:

Increased the LLCV to 20 from 19, due to amount of time it took to overfill CP3 (1476 seconds).

chandra.romel@LIGO.ORG - 16:30, Wednesday 21 December 2016 (32821)
I recommend raising CP3 even more to last through the break. Exhaust pressure will register a value above zero when it's boreline overfilling.
Displaying report 1-1 of 1.