Displaying report 1-1 of 1.
Reports until 10:41, Tuesday 03 January 2017
H1 PSL
corey.gray@LIGO.ORG - posted 10:41, Tuesday 03 January 2017 (32917)
PSL Weekly Report


Laser Status:
SysStat is good
Front End Power is 34.18W (should be around 30 W)
Front End Watch is GREEN
HPO Watch is GREEN
PMC:
It has been locked 0.0 days, 20.0 hr 28.0 minutes (should be days/weeks)
Reflected power is 13.29Watts and PowerSum = 80.42Watts.

---->There is a note that this PMC Reflected power is high, but that was for O1 (for O2 we are fine).

FSS:
It has been locked for 0.0 days 1.0 hr and 5.0 min (should be days/weeks)
TPD[V] = 3.673V (min 0.9V)

ISS:
The diffracted power is around 2.6% (should be 3-5%)
Last saturation event was 0.0 days 0.0 hours and 0.0 minutes ago (should be days/weeks)

THis closes FAMIS 7419.

Displaying report 1-1 of 1.