LHO VE
chandra.romel@LIGO.ORG - posted 12:05, Monday 09 January 2017 - last comment - 22:33, Monday 09 January 2017(33105)
IP5 controller replaced

Replaced IP5 controller in mech. room with GAMMA LPC (+) controller (WP 6421). One of two channels of the old controller was faulting "over temperature," a typical error/failure mode on the Varian controllers. LVEA pressures rose slightly due to reduced pumping speed (see aLOG 33070) over weekend. We will update CDS signal cabling (right now pump appears to be in error, but it is actually fully functioning) and reboot Beckhoff at next opportune maintenance period.

Comments related to this report
gerardo.moreno@LIGO.ORG - 22:33, Monday 09 January 2017 (33118)VE

Installed new cable for IP5, after removing old cables/connectors for previous version of controller.

Note: New controller is using the connection to pump A of old system, same calibration as other similar GAMMA (LPC +) controllers.