WP 6428
This morning the CO2 Laser Interlock Chassis D1200745 was replaced. This work is part of the ongoing investigation of the TCS flow sensor alarms/glitches. See alog 32776. Verified trip point for the temperature shutoff of the laser on new chassis. Steps taken to replace chassis:
1. Turn off laser at key
2. Turn off chiller in mechanical room
3. Turn off chassis and unplug all connections
4. Install new chassis, rocker switch set to CW (OPS) position
5. Redo connections to chassis & power on
6. Turn on chiller
7. Check there are no red lights on chassis front panel
8. Key on laser and press gate button
9. Check laser power actually powers on (MEDM screen).
Old Chassis S1302125
New Chassis S1302122
Fil, Richard, Nutsinee, Alastair
This swap unfortunately didn't seem to fix the problem. TCSY flowrate continues to glitch.