WP 6430
Chandra reported issues with Y2-8 ion pump, see alog entry. Gerardo verified controller had not tripped off. Following error messages were displayed on the controller:
1. Controller found with "Error 02"
2. After restarting controller, "Error 10", excessive arcing condition has been detected
With HV cable disconnected at power supply, HV output was enabled without no issues. Disconnected HV cable at both ends, and tested with the HiPOT tester. Test failed at ~500V. Marc tested the cable with the fieldfox N9912A, and will post some data.
Fil, Gerard, Marc
We compared scans from 8-2-2016 to today.
The short appears to be in approximately the same location as the previous short.