Starting CP3 fill. LLCV enabled. LLCV set to manual control. LLCV set to 50% open. Fill completed in 16 seconds. LLCV set back to 14.0% open. Starting CP4 fill. TC A error. LLCV enabled. LLCV set to manual control. LLCV set to 70% open. Fill completed in 32 seconds. LLCV set back to 35.0% open.
Lowered CP4 LLCV from 35% to 34% open because TCs are reading around -30degC.
CP3 filled fast but TCs are reading temps as expected.