Reports until 12:10, Monday 16 January 2017
LHO VE
logbook/robot/script0.cds.ligo-wa.caltech.edu@LIGO.ORG - posted 12:10, Monday 16 January 2017 - last comment - 16:09, Monday 16 January 2017(33342)
CP3, CP4 Autofill 2017_01_16
Starting CP3 fill. LLCV enabled. LLCV set to manual control. LLCV set to 50% open. Fill completed in 420 seconds. LLCV set back to 14.0% open.
Starting CP4 fill. LLCV enabled. LLCV set to manual control. LLCV set to 70% open. Fill completed in 60 seconds. TC A did not register fill. LLCV set back to 34.0% open.
Images attached to this report
Comments related to this report
gerardo.moreno@LIGO.ORG - 16:09, Monday 16 January 2017 (33346)

Remote logged in and changed the setting for CP3 LLVC due to the amount of time it took to overfill, I did single unit steps, started around 19:41 utc, changed it to 15 from 14, watched it for 10 minutes, but no increase on pressure and temperature remained the same.

A second change at 19:53 utc, to 16 from 15, no noticeable change on temperature, but pressure went up to 0.1 psig.

A third change at 21:04 utc since themocouple temperature was still hovering around -10 oF, setting changed to 17 from 16, pressure went up to 0.3 psig, and temperature started moving, on the course of 4 hours it went down to -19 oF from -10.

I wiill leave it at this setting (17%) and re-evaluate tomorrow, since ambient temperature suppose to go up.