Note: There was a PSL trip last night & also ISS electronics work this morning for Maintenance.
Laser Status:
SysStat is good
Front End Power is 34.11W (should be around 30 W)
Front End Watch is GREEN
HPO Watch is GREEN
PMC:
It has been locked 0.0 days, 0.0 hr 10.0 minutes (should be days/weeks)
Reflected power is 17.05Watts and PowerSum = 76.71Watts.
FSS:
It has been locked for 0.0 days 0.0 hr and 10.0 min (should be days/weeks)
TPD[V] = 1.157V (min 0.9V)
ISS:
The diffracted power is around 4.2% (should be 3-5%)
Last saturation event was 0.0 days 0.0 hours and 9.0 minutes ago (should be days/weeks)
Possible Issues:
PMC reflected power is "noted" as being high in our PSL Report script. Attached is a 1-month trend of PMC Refl
This closes FAMIS 7421.