Reports until 16:17, Wednesday 18 January 2017
LHO General
patrick.thomas@LIGO.ORG - posted 16:17, Wednesday 18 January 2017 - last comment - 17:29, Wednesday 18 January 2017(33413)
no FMCS alarm levels recorded in burt snapshots since Aug 25 2016, no alarm levels currently set
Discovered while trying to determine why it appears that there are no alarm levels currently set for the FMCS channels.

It appears the last burt snapshot for the FMCS IOC that contains an alarm level is /ligo/cds/lho/h0/burt/2016/08/25/11:00/h0fmcs.snap.

Found by running:
cd /ligo/cds/lho/h0/burt/2016
grep -r HIGH */*/*/h0fmcs.snap

The last file listed is 08/25/11:00/h0fmcs.snap.

The autoBurt.req file in /ligo/lho/h0/target/h0fmcs has a modification date of Aug 25 11:34. It does not contain any alarm levels.
Comments related to this report
patrick.thomas@LIGO.ORG - 16:22, Wednesday 18 January 2017 (33414)
I burtrestored the IOC to the snapshot taken on Aug. 25 2016 11:00 and wrote 12.0 to H0:FMC-MILLIAMP_MAXVAL.

I'm not certain what I did to lose the autoBurt.req file that has the alarm levels in it. I will work to restore it.
patrick.thomas@LIGO.ORG - 16:45, Wednesday 18 January 2017 (33415)
It looks like the last snapshot file with a severity set is /ligo/cds/lho/h0/burt/2016/04/13/12:00/h0fmcs.snap (April 13 2016 12:00).
patrick.thomas@LIGO.ORG - 17:29, Wednesday 18 January 2017 (33417)
I have updated the autoBurt.req file in /ligo/lho/h0/target/h0fmcs. This one has the alarm level and severity channels in it. I have also added it to the projects svn repository in trunk/epics/iocs/fmcs.

I have burtrestored the IOC using the snapshot from April 13 2016 (/ligo/cds/lho/h0/burt/2016/04/13/12:00/h0fmcs.snap).