Reports until 12:10, Monday 23 January 2017
LHO VE
logbook/robot/script0.cds.ligo-wa.caltech.edu@LIGO.ORG - posted 12:10, Monday 23 January 2017 - last comment - 14:27, Monday 23 January 2017(33538)
CP3, CP4 Autofill 2017_01_23
Starting CP3 fill. LLCV enabled. LLCV set to manual control. LLCV set to 50% open. Fill completed in 17 seconds. TC B did not register fill. LLCV set back to 17.0% open.
Starting CP4 fill. LLCV enabled. LLCV set to manual control. LLCV set to 70% open. Fill completed in 2275 seconds. LLCV set back to 36.0% open.
Images attached to this report
Comments related to this report
chandra.romel@LIGO.ORG - 14:27, Monday 23 January 2017 (33542)

Raised CP4 LLCV from 36% to 37% open.