Arnaud had sent along some measurements on HAM1, trying to measure the transmissibility from the HAM1 stack to the RM suspensions. I ran these measurments this morning, during the maintenance period. It looks like the LLO and LHO RM damping designs might differ a bit, maybe the LLO damping filter is a bit more aggressive? First image is just the RM1 L plant measurement, it was taken with the RMs both in DAMPED (unaligned) and HEPI isolated (SC was off, but HAM1 uses high UGF, blended (L4C/IPS) loops). The second image is the HEPI X to RM1 L transfer function, which was taken with HEPI in READY and the RMs still DAMPED. Both templates are saved in the SVN in H1/ HAM1/Data/Transfer_Functions/Measurements/Undamped folder.
EDIT: LLO's measurements are the black refs, my measurements are the red. Arnaud reports that the difference is due to an old damp filter design on the RMs, LLO now matches us.
Jim, I attached the results in alog 31179.