john.oh@LIGO.ORG - posted 03:56, Friday 27 January 2017 (33671)
DQ Shift Report for 23-25 January
Overall duty cycle: 72.14 % (3 days)
Inspiral range around 65-70Mpc - range.
Degradation in inspiral range (by ~5-10Mpc) coincides with noise excess in 50-200Hz appearing in the spectrogram for which the reason is unclear.
It could be correlated with rubbing optics, but degradation of inspiral range is less severe than typical rubbing and the frequency band of the noise is narrower than by typical rubbing.
But still there exists similarity between the rubbing effect and the noise excess.
Hveto vetoed mostly low-frequency glitches (≲10Hz), but not in higher frequency range.
Typical range decrease from 13:00 UTC due to anthropogenic ground motion at Hanford