Reports until 16:30, Tuesday 31 January 2017
H1 CAL (CAL)
travis.sadecki@LIGO.ORG - posted 16:30, Tuesday 31 January 2017 (33783)
End X PCal Calibration

Yuki, Heather, Travis, Rick

Results of the latest End X PCal calibration can be found on the DCC at T1500129 and T1500622.  Although the baseline is rather short since the Working Standard was replaced before the last 2 measurements, the variation in the PDs is less than the 1% required by Calibration.  Results have been reviewed by Rick, myself, and Yuki and the only standout we noted was the power imbalance of the inner and outer beams, which varied by 1.2% from the mean.  We will continue to monitor this imbalance in further measurements and address it if it persists.