Laser Status:
SysStat is good
Front End Power is 34.05W (should be around 30 W)
HPO Output Power is 164.3W
Front End Watch is GREEN
HPO Watch is GREEN
PMC:
It has been locked 7 days, 21 hr 17 minutes (should be days/weeks)
Reflected power = 14.81Watts
Transmitted power = 65.69Watts
PowerSum = 80.5Watts.
FSS:
It has been locked for 0 days 14 hr and 11 min (should be days/weeks)
TPD[V] = 1.753V (min 0.9V)
ISS:
The diffracted power is around 2.0% (should be 3-5%)
Last saturation event was 0 days 14 hours and 12 minutes ago (should be days/weeks)
Possible Issues:
ISS diffracted power is Low