Reports until 16:14, Friday 10 February 2017
H1 General
jeffrey.bartlett@LIGO.ORG - posted 16:14, Friday 10 February 2017 (34051)
Ops Day Shift Summary

Ops Shift Log: 02/10/2017, Day Shift 16:00 – 00:00 (08:00 - 16:00) Time - UTC (PT)

State of H1: Locked at NOMINAL_LOW_NOISE, 30.3 W of power and range of 68.0 Mpc.      
Intent Bit: Observing for the past 7 hours.
Support: Sheila, Terra, Jim B.,
Incoming Operator: Ed

 

Shift Summary: Encountered relocking problems with the OMC LSC_I filter bank, which Sheila resolved. During and after locking worked with Terra to setup Ops StripTools to monitor the two new PI_Modes (Mode19 & Mode20). The PI2_monitoring.stp has been reloaded on Nuc6. Had to damp PI_Modes 26, 27, & 28 several times during the shift. Also took advantage of LLO not having fully recovered from this morning’s lockloss to run the A2L script. By the end of the shift, the A2L Pitch was elevated again.

Sheila & Terra implemented new filters to damp an additional PI_Mode that was implicated in the last two lockloss, (see aLOG #34046 and WP #6474). Dropped out of Observing for 2 minutes to clear SDF differences.  

The wind is starting to come up again, now at Moderate to Strong Breeze (13 to 30 mph); End-Y microseism is up sharply.     

 

Activity Log: Time - UTC (PT)
16:00 (08:00) Take over from Jim
16:00 (08:00) Relocking after EQ
16:58 (08:58) Relocked at NOMINAL_LOW_NOISE
17:19 (09:19) Damp PI_Mode28
17:21 (09:21) Damp PI_Mode27
17:22 (09:22) Terra – Cleared SDFs from adding PI_Mode19 and PI_Mode20
17:23 (09:23) Set Intention Bit to Observing
17:33 (09:33) Damp PI_Mode27
17:43 (09:43) In Commissioning/Corrective Maintenance to run A2L Script (LLO Down)
17:45 (09:45) Jeff K – Going to Roof to take pictures
17:50 (09:50) Jeff K – Off the roof
17:54 (09:45) Set Intent Bit to Observing
18:08 (10:08) Platt Electrical – On site for delivery – Notified Richard
18:15 (10:15) Kyle – Going to Mid-Y for vacuum work (WP #6472)
18:33 (10:33) Betsy – Going into the Optics lab
18:35 (10:35) Damped PI_Mode28
18:55 (10:55) Betsy – Out of the Optics lab
18:57 (10:57) Damped PI_Mode26
19:14 (11:14) Karen – Cleaning in the Optics Lab
19:45 (11:45) Alarm on PT245 (CP4 at Mid-Y) – Notified VAC group
19:55 (11:55) Kyle – Leaving Mid-Y
19:57 (11:57) Damped PI_Mode27
19:59 (11:59) Karen – Finished in the Optics Lab
20:35 (12:35) Kyle – Going to Mid-Y for vacuum work (WP #6472)
22:32 (14:32) Kyle – Back from Mid-Y
23:00 (15:00) Drop out of Observing – (SDF) Load of new filters for new PI Mode (WP #6474)
23:01 (15:01) Back to Observing
23:01 (15:01) Drop out of Observing (SDF)
23:02 (15:02) Back to Observing
00:00 (16:00) Turn over to Ed