Kyle noted y-arm pressure rise last night. This morning I found the controller for Y2-8 ion pump faulted with error 02. Rebooted and came back at 350 V. Plugged pump into second channel and read 700 V at 0.5 A.
Marc and Richard tested the HV cable and concluded the cable connector on pump side is faulty. Marc will comment with his results.
End Y
Y2-8
End Y
Conclusion
It looks as if the ion pump end of the cable is shorted in or near the connector at Y2-8. The cable near the end of the connector does not have strain relief and is slightly deformed. Manipulating the cable near the connector did not change the cable VSWR trace, but it did change the measured ohms slightly. More can be learned from disassembly of the connector housing without damaging the connector. This should be investigated and the connector repaired / replaced next maintenance window.