I have produced filters for offline calibration of Hanford data starting at GPS time 1169326080. The filters can be found in the calibration SVN at this location:
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/GDSFilters/H1DCS_1169326080.npz
For information on the associated change in calibration, see:
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=33585
For suggested command line options to use when calibrating this data, see:
https://wiki.ligo.org/Calibration/GDSCalibrationConfigurationsO2
The filters were produced using this Matlab script in SVN revision 4251:
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/Scripts/TDfilters/H1_run_td_filters_1169326080.m
The parameters files used (all in revision 4251) were:
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/params/2017-01-24/modelparams_H1_2017-01-24.conf
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/params/2017-01-24/H1_TDparams_1169326080.conf
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/Scripts/CAL_EPICS/D20170124_H1_CAL_EPICS_VALUES.m
Several plots are attached. The first four (png files) are spectrum comparisons between CALCS, GDS, and DCS. GDS and DCS agree to the expected level. Kappas were applied in both the GDS plots and the DCS plots with a coherence uncertainty threshold of 0.4%.